Home  Cardiology  EKG Machines  GE / Marquette MAC 500 EKG Machine

GE / Marquette MAC 500 EKG Machine

GE / Marquette MAC 500 EKG Machine
Item #none
AvailabilityContact us for availability.
Please call for a price.
This item comes New.
Send Page to FriendPrint this page

View Complete SpecificationsQuick Specs!

Cost-effective Electrocardiography
The MAC 500 resting ECG system is a portable twelve lead electrocardiograph, designed for acquisition, with or without analysis. The system can be configured as a very basic electrocardiograph without analysis for those facilities needing minimal ECG requirements-or be configured to incorporate the proven Marquette 12SLTM ECG analysis program. By adding the 12SL program into the system you will create an electrocardiograph with the best "second opinion" analysis and measurements program available. If your facility would only…

Cost-effective Electrocardiography
The MAC 500 resting ECG system is a portable twelve lead electrocardiograph, designed for acquisition, with or without analysis. The system can be configured as a very basic electrocardiograph without analysis for those facilities needing minimal ECG requirements-or be configured to incorporate the proven Marquette 12SLTM ECG analysis program. By adding the 12SL program into the system you will create an electrocardiograph with the best "second opinion" analysis and measurements program available. If your facility would only require measurements, we offer a version which includes 12SL measurements only. Whatever the situation, the MAC 500 resting ECG system most certainly can be a very cost-effective addition to compliment your resting ECG requirements.


Marquette MAC® 500 Resting ECG System Specifications Options
A carrying case with shoulder belt is available.

Technical specifications
1/3 channel resting ECG system. ECG acquisition and direct recording of waveforms and alphanumeric characters to thermal sensitive paper.

Processing
Acquisition: 12-leads (simultaneously)
Signal input: Type CF according to IEC
Preacquisition: Provides 10 seconds of instantaneous ECG acquisition and printing of any selected report format
Digital sampling rate: 1000 samples/second/channel
Dynamic range: Differential signals for AC voltage ±10 mV
Superimposed DC voltage (Polarization voltage) ±600 mV
Resolution: 5 µV
Frequency response: -3 dB @ 0.05 - 150 Hz (IEC/AHA)
Common-mode rejection: With AC-Filter switched on >140dB
Input impedance: For common mode signals referred to RL >50 M Ohm for up to 60 Hz, defibrillation protected
Pacemaker detection sensitivity: Duration > 0.1 ms, amplitude > 5 mV

Display
Type: Alphanumeric LCD with 2 x 16 characters
Contrast adjustable
Displayed data: Selected lead group, operation mode, lead/patient cable off, heart rate, data acquisition, AC filter, muscle filter, gain, paper empty

Writer
Type: Thermal array print head
Resolution: Vertically 8 dots/mm, horizontally 12.5 µm at 25 mm/s
Paper width: 3.5 in (90 mm)/80 mm writing width
Writer speed: 5, 25, 50 mm/s
Paper type: roll

Keyboard
Type: Membrane keypad/17 keys

Special functions
Anti Drift System (ADS): Automatic compensation of baseline fluctuations caused by polarization voltage fluctuations at the electrodes
Electrode monitoring: Audible and visual indication on the LCD of disconnected electrodes or line break. Each single electrode is monitored.
Copy: In the automatic 12 lead mode, after ECG recording, copies of the ECG can be printed
Test: Automatic performance test upon power up, including verification of the signal path starting at the signal input

Power supply
Rated voltage: 110 - 240 V
Frequency: 49 to 65 Hz
Power consumption: 28 VA max
Battery type: lead acid, 12 V, 1.3 Ah

Physical specifications
Height: 3.2 in (80 mm)
Width: 11.5 in (290 mm)
Depth: 7.9 in (200 mm)
Weight: 4.4 lbs. (2 kg) w/battery

View Complete Specifications