The B125 patient monitor from GE Healthcare delivers clinical
performance you can trust. It is simple and intuitive for busy
caregivers, flexible for diverse care areas and acuity needs, and
reliable for challenging work environments.
Parameter excellence
The B125 Monitor is designed with advanced parameter
technologies for accurate and reliable patient monitoring:
- • EK-Pro arrhythmia analysis
- • DINAMAP SuperSTAT non-invasive blood pressure
- • TruSignal enhanced SpO2 saturation monitoring. Also available with Nellcor® OxiMax® SpO2 or Masimo® SET® SpO2 algorithms
- • GE EtCO2 sidestream measurement
- • Comprehensive package for neonatal *1 measurements
Flexible for diverse care areas & acuity
needs
Designed with advanced technologies to support connectivity
& operations across hospital enterprise.
- • Connectivity to GE CARESCAPE networks and Mobile
Care Web Viewer
- • Flexibility to share parameter modules and accessories
across other GE monitors
- • Advanced applications such as ST Segment and Full
Arrhythmia
Simple & intuitive user interface
The B125 Monitor makes it easy to acquire, monitor and
interpret patient data to support timely decision-making:
- • Seven pre-configured workflow settings
- • Choice of numerical or continuous waveform monitoring
- • Big numeric view capability
- • Capacitive touch screen
Robust & reliable
Its rugged & robust design stands up to harsh environments
and the everyday wear-and-tear of busy care areas.
- • Upgraded cyber security
- • Stable performance in tough environmental conditions
- • 75 cm height drop test for main unit
- • Three hour battery
*1 Impedance respiration is intended for use with only adult and pediatric patients in United States, Argentina, Guam, Puerto Rico, Saint Croix and Saint Thomas.
CO2 measurement through E-miniC Module is intended for use with patients weighing over 5kg (11 lb) only.
Display
- Size: 12.1 in (diagonal)
- Resolution: 1280 x 800 pixels (WVGA)
- Number of waveforms: Up to 6
- Display layout and colors: User-configurable
- Controls: Capacitive Touch screen & Trim
Knob control and hard keys
(standard)
Parameters and modules
ECG
Leads available:
3-lead configuration: I, II, III
5-lead configuration: I, II, III, aVR,
aVL, aVF and V
Sweep speed: 12.5, 25 or 50 mm/s
Gain range: 0.5x, 1x, 2x and 4x
Heart rate accuracy: 30 to 300 bpm, ±5% or ±5 bpm,
whichever is greater
Bandwidth
50/60 Hz power supply:
Monitor: 0.5 to 40 Hz
ST:
0.05 to 40 Hz
Diagnostic: 0.05 to 145 Hz
Pacemaker detection:
Range: 2 to 700 mV
Pulse width: 0.5 to 2 ms
Arrhythmia analysis:
Asystole, V Fib / V Tach, V Tach, VT>2
R on T, V Brady, Couplet, Bigeminy,
Accelerated Ventricular arrhythmia,
Multifocal PVCs, A Fib, Missing beat,
Pause, Tachy, Brady, Trigeminy
ST segment analysis:
Numeric range: -9 to +9 mm
(-0.9 to +0.9 mV)
Accuracy: ±0.2 mm or ±10%,
whichever is greater, within the
measurement range of -8 to 8 mm
Numeric resolution: 0.1 mm
(0.01 mV)
ST Trends: Up to 168h
Impedance respiration *3
Range:
Adult/pediatric: 4 to 120 resp/min
Neonate*3: 4 to 180 resp/min
Accuracy: ±5% or ±5 resp/min, whichever is
greater
Gain: range 0.1 to 5 cm/Ohm
SpO2
TruSignal SpO
2
Measurement range
Pulse oximetry: 1 to 100%
Pulse rate: 30 to 250 bpm
Measurement accuracy:
Saturation Without motion-adult/pediatric
Finger sensor: 70 to 100% ±2%
Without motion-neonate*3:
70 to 100% ±3%
With motion-adult/pediatric/
neonate*3: 70 to 100% ±3%
Low perfusion-adult/pediatric:
70 to 100% ±3%
(1~69% unspecified)
Pulse Rate Without motion: ±2 bpm
(Adult/Pediatric/Neonatal*3)
*2 Refer to B125 User's Manual for more information.
*3 Impedance respiration is intended for use with only adult and pediatric patients in
United States, Argentina, Guam, Puerto Rico, Saint Croix and Saint Thomas.
CO
2 measurement through E-miniC Module is intended for use with patients weighing
over 5kg (11 lb) only.